IDQ ID281
This manual contains the instructions for the ID Quantique (IDQ) SNSPD ID281.
There are four detectors: two in the visible range, and two in the near-infrared range.
Cooldown (new system)
The system needs to be pumped down before starting the cooldown process. We do not own a vacuum pump at the moment, but we typically borrow it from other groups.
- Borrow the vacuum pump
- Make sure that the He static pressure on the compressor is higher than 1.65 MPa;
- From the web interface on the second PC (accessible at 169.254.28.1:8080), verify that the system is in the Starting status, waiting for the user to start the sequence.
- Press the start button on the web UI.
- Follow the instructions on the web UI as they guide you through the different steps. Roughly, you will:
- Slowly open the manual venting valve
- Connect the vacuum pump
- Switch on the pump power supply
- Switch on the turbo pumping by pressing the main button (⏼)
- Scroll through the parameters until you see the pressure value (parameter 340: Pressure)
- When pressure reaches
hPa, turn on the He compressor switch to start cooling the system
- Once you get the message Fully close the manual venting valve, and then confirm, do so.
- Stop the pump by pressing the main button (⏼), and wait for it to stop rotating.
- Disconnect the vacuum pump.
- Cover the valve with the blank flange, centering ring, and cast clamp.
The compressor cooling phase takes ~8 hours from room temperature.
Operating the new software
The new software is based on a web interface, that can currently be accessed at the address 169.254.28.1:8080. The connection to the ID281 Control Box occurs via Ethernet cable.
Updating the calibration files
This calibration should be performed when the system is at room temperature.
- Download the XXXXX software and connect via USB to the ID281 Control box
- For a given input, disconnect from profibus
- Click on Add/replace sensor
- Click on Custom curve and select the calibration file
- Repeat for the other inputs, if necessary
- Finish by allowing the connection to profibus
- Proceed to cooldown the system
Manuals
Manuals, installation certificates, and tutorials are available in the O: drive "User manuals" folder.